Calibration Test and PDF Report

Opening the test panel

Click Calibration Test in the top bar. The panel slides in from the right.

Test information

FieldDescription
OperatorPerson performing the test
Reference deviceReference gauge or calibrator used
ToleranceAllowed error (% of full scale), for example 0.25
SamplesNumber of readings averaged at each point

Creating test points

  • Generate 5 points: Creates points at 0%, 25%, 50%, 75% and 100% of the measuring range.
  • Add point: Adds an empty row where you enter the reference value yourself.
  • Trash can: Deletes all points.

Taking measurements

  1. Apply the pressure for the first point using the reference device.
  2. When the reading settles, click Measure on that row.
  3. The application takes the set number of readings and enters their average.
  4. Repeat for the other points.

Each row shows the measured value, the error and the error as % FS. Rows are marked PASS (green) or FAIL (red) against the tolerance. The largest error and the overall result appear below the table.

Exporting the report

  • PDF Report: Creates an A4 report with device information, the test table, the result, notes and a signature field.
  • CSV: Saves the test table as a file that opens in Excel.