Analog Output Calibration

Analog output points

The 4mA Output and 20mA Output fields set which pressure values the analog output corresponds to.

Example: entering 0 and 10 gives 4 mA at 0 bar and 20 mA at 10 bar.

Calibration steps

  1. Connect an ammeter to the transmitter's current output.
  2. Turn on the Calibration of analog output switch and confirm.
  3. Select 4mA Cal. The device forces its output to 4 mA.
  4. Watching the ammeter, use the Trim buttons to bring the output to exactly 4.000 mA:
    • « / » : 10 steps
    • ‹ / › : 1 step
  5. Select 20mA Cal and repeat for 20.000 mA.
  6. Turn the calibration switch off. The device returns to normal output.
  7. Click Save.

Note: If the application is closed while calibration is on, it sends a command to release the forced output. Even so, always finish calibration by turning the switch off.